Glossary of Pressure Terms
Absolute Pressure: pressure measured relative to full vacuum
- Referred to as pounds per square inch absolute (PSIA)
Accuracy: combined error of linearity, hysteresis, and repeatability
- Setra uses the root sum of the squares (RSS) method
Ambient Conditions: the conditions (pressure, temperature, etc.) of the medium surrounding the case of the transducer
Atmospheric Pressure: pressure of the atmosphere at Earth's surface
- At sea level this pressure is not constant (it varies with the weather), but averages around 100 kilopascals (kP)
- Ambient pressure may, in other circumstances, be measured in pounds per square inch (psi) or in atmospheres (atm)
- One atmosphere is also approximately the ambient pressure at sea level and is equal to 14.7 psi
BAR: Unit of pressure (or stress)
- 1 bar = 750.07 mm of mercury at 0°C, lat. 45°
Barometric Pressure: atmospheric pressure, often measured in millibars, in Hg (inches of mercury), or hectopascals
Bourdon Tube: an early, mechanical pressure gauge consisting of a flattened tube that tends to straighten under internal pressure
- Today usually used with a potentiometer to produce an electrical output
Burst Pressure: the maximum pressure that may be applied to the positive pressure port without rupturing the sensing element
Capacitive Sensing: detection and measurement of pressure through the change in voltage across a capacitor, one plate of which is a diaphragm which deflects slightly which changes in applied pressure
Compound Pressure: pressure measured from full vacuum (-14.7 PSIV) to gauge pressure, referencing atmosphere
Differential Pressure: pressure measured relative to a reference pressure
- Referred to as pounds per square inch differential (PSID)
Excitation: the external electrical voltage and/or current applied to a transducer for its proper operation
FS (Full Span or Full Scale): the range of measured values over which a transducer is intended to measure, specified by the upper and lower limits
- Ex: 0 to 100 PSIG
- FS is 100 PSIG to 5 VDC
- FS is 5 VDC
- 800-100 MB FS is 300 MB
Gage Pressure: pressure measured relative to ambient atmospheric pressure
- Referred to as pounds per square inch gauge (PSIG)

Hysteresis: the maximum difference in output, at any measured value within the specified range, when the value is approached first when increasing and then decreasing pressure

Linearity: the maximum deviation of any calibration point on a specified straight line, during any one calibration cycle
- Best Fit Straight Line (BFSL) Method: relationship of a calibration curve to a specified straight line

- End Point Method: relationship of a calibration curve to a specified straight line through its end points

- Terminal Method: relationship fo a calibration curve to a specified straight line with end points at zero and full scale

Long Term Stability: the ability of a transducer to reproduce output readings obtained during its original calibration at room conditions for a specified period of time, usually one year
Manometer: an early instrument for measuring pressure
- Originally a U-shaped tube containing liquid (water, oil, or mercury), one limb opening to the gas volume to be measured, the other closed or connected to a registering or recording instrument
- Modern versions utilize diaphragms, bellows, or other devices for sensing relative pressures
Millibar (mbar): unit of pressure generally used in barometric measurements
- 1 mbar = 100 N/m2, or 1~ dyn/cm2
Newton (N): the unit of force in the International System of Units (SI)
- The force required to impart an acceleration of 1m/sec2 to a mass of 1 kg
Pascal (Pa): the standard unit of pressure (or stress) in the SI system
- Equal to 1 newton per square meter (1 N/m2)
P/I: term common to process industries meaning pressure-in/current-out
- 3-15 PSIG Input to 4 to 20 mA DC Output
Position Effect: gravity effect on the output of a transducer when mounted with the diaphragm in different orientation than when it was calibrated
- Effect is usually minimal and can be zeroed out with the zero potentiometer (on units with a zero pot) with no effect on accuracy
Pressure Transducer: an electromechanical device for translating fluid pressure valves into voltages across a high-impedance (5k ohms or greater) load
Pressure Transmitter: an electromechanical device for translating fluid pressure values into currents (generally 4 to 20 mA) in a low-impedance load
Proof Pressure: the maximum pressure that may be applied without changing performance beyond specifications (typically 0.5% FS zero shift)
- If the proof pressure is constantly applied over time, there may be an offset in output attributed to it
PSIA: pounds per square inch absolute
PSIV: pounds per square inch vacuum
Range: the spread between the maximum and minimum pressure between which the transducer has been designed to operate
Repeatability: the ability to reproduce output readings when the same pressure value is applied consecutively, under the same conditions, and in the same direction

Response Time: the length of time required for the output to rise to a specified percentage of its final value as a result of a step change in pressure
RSS Accuracy: the root sum of the squares
- Combined error of linearity, hysteresis, and repeatability (at a constant temperature)

Setraceram™: a patented ceramic/glass sensing element used by Setra in barometric transducers and instruments requiring the highest degree of measurement precision and stability
Span: the algebraic difference between the limits of the range
- Ex: 0.1 to 5.1 Volts DC
- Span is 5 VDC
- Sometimes used to designate full scale output (ex: 5 VDC)
Thermal Error: the maximum change in output, at any pressure value within the specified range, when the temperature is changed from room temperature to specified temperature extremes
Thermal Zero Shift: the zero shift due to changes of the ambient temperature from room temperature to the specified limits of the operating temperature range
Torr: a unit of low pressure equal to a head of 1 mm of mercury or 133.3 N/m2
Vacuum: generally refers to pressures between 0 and atmospheric
- Often measured in 0-30 Hg Vacuum
- Referred to as pounds per square inch vacuum (PSIV)